Stockert Heater Cooler System 3T (2018-10-28)
- Starting date:
- October 28, 2018
- Posting date:
- November 15, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68236
Affected products
Stockert Heater Cooler System 3T
Reason
LivaNova is providing updated instructions to monitor the concentration of hydrogen peroxide in the water circuit of the 1T and 3T Heater-Cooler devices, to verify that sufficient concentration of hydrogen peroxide is present to limit microbial growth, and to adjust the concentration of hydrogen peroxide if it drops below 100 ppm. This regimen enhances the water maintenance regimen outlined in Section 6.4 of the operating instructions.
Affected products
Stockert Heater Cooler System 3T
Lot or serial number
More than 100 numbers, contact manufacturer
Model or catalog number
- 16-02-81
- 16-02-82
- 16-02-85
Companies
- Manufacturer
-
Livanova Deutschland GmbH
Lindberghstrasse 25
Munich
80939
GERMANY