Health product recall

Stockert Heater Cooler System 3T (2018-10-28)

Starting date:
October 28, 2018
Posting date:
November 15, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68236

Affected products

Stockert Heater Cooler System 3T

Reason

LivaNova is providing updated instructions to monitor the concentration of hydrogen peroxide in the water circuit of the 1T and 3T Heater-Cooler devices, to verify that sufficient concentration of hydrogen peroxide is present to limit microbial growth, and to adjust the concentration of hydrogen peroxide if it drops below 100 ppm. This regimen enhances the water maintenance regimen outlined in Section 6.4 of the operating instructions.

Affected products

Stockert Heater Cooler System 3T

Lot or serial number

More than 100 numbers, contact manufacturer

Model or catalog number
  • 16-02-81
  • 16-02-82
  • 16-02-85