This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Stimpod NMS450 Nerve Stimulator (2015-12-15)

Starting date:
December 15, 2015
Posting date:
June 30, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59100

Affected products

A. Stimpod NMS450 Nerve Stimulator

Reason

In isolated incidences (less than 0.003%) while using the device in NMBA monitoring mode, electrodes used may deteriorate to a level where current densities may be large enough to cause superficial burns on the skin of the patient.

Affected products

A. Stimpod NMS450 Nerve Stimulator

Lot or serial number

All lots

Model or catalog number

XT-41000-NA

XT-41001-NA

Companies
Manufacturer

Xavant Technology (PTY) LTD.

Unit 102 Tannery Industrial Park, 309 Derdepoort Road

Silverton

0184