This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

STERRAD Product Recall (2014-12-08)

Starting date:
December 8, 2014
Posting date:
December 12, 2014
Type of communication:
Drug Recall
Subcategory:
Drugs
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Product Safety
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-42919

Recalled Products

  1. STERRAD 50 Cassette
  2. STERRAD 100 Cassette LIQ 58%
  3. STERRAD 100NX Cassette
  4. STERRAD NX Cassette

Reason

The leak indicator on the STERRAD System cassettes packaging assembly may not function properly after exposure to hydrogen peroxide. The leak indicator is designed to provide a visual color indication (it turns from yellow to red) in the event hydrogen peroxide has leaked form the cassettes, however, it may be unstable and may not remain red for the duration of the products' shelf-life.

Depth of distribution

Hospitals

Affected products

A. STERRAD 50 Cassette

DIN, NPN, DIN-HIM

DIN 02250365

Dosage form

Liquid

Strength

Hydrogen Peroxide 58%

Lot or serial number

All Lots

Companies
Recalling Firm

Johnson & Johnson Medical Products

200 Whitehall Drive

Markham

L3R 0T5

Ontario

CANADA

Marketing Authorization Holder

Johnson & Johnson Medical Products

200 Whitehall Drive

Markham

L3R 0T5

Ontario

CANADA



B. STERRAD 100 Cassette LIQ 58%

DIN, NPN, DIN-HIM

DIN 01980475

Dosage form

Liquid

Strength

Hydrogen Peroxide 58%

Lot or serial number

All Lots

Companies
Recalling Firm

Johnson & Johnson Medical Products

200 Whitehall Drive

Markham

L3R 0T5

Ontario

CANADA

Marketing Authorization Holder

Johnson & Johnson Medical Products

200 Whitehall Drive

Markham

L3R 0T5

Ontario

CANADA



C. STERRAD 100NX Cassette

DIN, NPN, DIN-HIM

DIN 02296667

Dosage form

Liquid

Strength

Hydrogen Peroxide 58%

Lot or serial number

All Lots

Companies
Recalling Firm

Johnson & Johnson Medical Products

200 Whitehall Drive

Markham

L3R 0T5

Ontario

CANADA

Marketing Authorization Holder

Johnson & Johnson Medical Products

200 Whitehall Drive

Markham

L3R 0T5

Ontario

CANADA



D. STERRAD NX Cassette

DIN, NPN, DIN-HIM

DIN 02253194

Dosage form

Liquid

Strength

Hydrogen Peroxide 58%

Lot or serial number

All Lots

Companies
Recalling Firm

Johnson & Johnson Medical Products

200 Whitehall Drive

Markham

L3R 0T5

Ontario

CANADA

Marketing Authorization Holder

Johnson & Johnson Medical Products

200 Whitehall Drive

Markham

L3R 0T5

Ontario

CANADA