Stellar 100 and Stellar 150 (2019-12-05)
- Starting date:
- December 5, 2019
- Posting date:
- December 20, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71937
Last updated: 2019-12-20
Affected Products
Stellar 100 and Stellar 150
Reason
Resmed received a confirmed case of an alarm buzzer not working under the following circumstances: the device has a failed electronic component and, the device is stored without AC power connected for more than 36 hours leading to full depletion of the battery and, the device powers on automatically when connected to AC power without pressing the power switch. The identified issue is a combination of software and a component failure that may occur in a small number of devices. If a device with is issue is used in the condition described above, the audible alarms may not operate properly.
Affected products
Stellar 100 and Stellar 150
Lot or serial number
20160123307 to 22171057208
Model or catalog number
24140
24157
24161
24163
24171