This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Statspin PM Centrifuge (SSMP) and RT12 Rotor Kit (2014-12-15)

Starting date:
December 15, 2014
Posting date:
January 7, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43149

Recalled Products

A) Statspin PM Centrifuge (SSMP)

B) RT12 Rotor Kit

Reason

Iris International has received reports of uncontained debris (part fragments) as a result of rotor breakage and lid failures on similar centrifuge models which use the RT12 rotor and shield.

If the RT12 rotor breaks and the centrifuge fails to contain debris, escaping debris may result in possible operator injury, direct exposure to biohazardous materials, and/or loss of sample. Rotor breakage may also result in damage to the centrifuge.

Affected products

A) Statspin PM Centrifuge (SSMP)

Lot or serial number

All lots.

Model or catalog number

M901-22

Companies
Manufacturer

Iris Diagnostics, a Division of Iris International Inc.

9172 Eton Avenue,

Chatsworth

91311

California

UNITED STATES



B) RT12 Rotor Kit

Lot or serial number

All lots.

Model or catalog number

Not applicable  

Companies
Manufacturer

Iris Diagnostics, a Division of Iris International Inc.

9172 Eton Avenue,

Chatsworth

91311

California

UNITED STATES