Health product recall

Stat Profile Prime® Plus Analyzer

Last updated

Summary

Product
Stat Profile Prime® Plus Analyzer
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalog number/
Stat Profile Prime® Plus Analyzer PP1218410C PP14210410C PP1918190C PP1C18130C PP12210030C PP12210080C PP1420880C PP15230050C PP16220090C PP1918340C PP1419210C PP14230260C PP12230670C PP18220170C PP18220090C PP1219230C PP1420180C PP1319060C PP15230030C PP18220130C PP18220180C PP19181902 PP1918321C PP1B18340C 57400 59508
Stat Profile Prime® Plus POC Analyzer PP31220460C PP33230230C 63368
Stat Profile Prime® Plus POC Analyzer (Remanufactured) Not applicable. 63369

Issue

Nova Biomedical became aware of a potential risk of reporting erroneous creatinine patient results associated with the initial start-up of a new bun/creatinine sensor cartridge. The potential risk is within the first 2-4 hours after cartridge installation that may not be identified by traditional quality control practices.

Recall start date: September 13, 2023

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Health products - Medical devices - Chemistry
Companies

Nova Biomedical Corporation

200 Prospect St., Waltham, Massachusetts, United States, 02454

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74618

Get notified