i-STAT Alinity Instrument with Analyzer Software (2017-12-21)
- Starting date:
- December 21, 2017
- Posting date:
- January 29, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65794
Affected products
i-STAT Alinity Instrument with Analyzer Software
Reason
This product action addresses two separate issues:
- Abbott Point of Care (APOC) has determined that an increased frequency of quality check failure code 90-02 when testing i-STAT cartridges with the i-STAT Alinity software version Osi05 (JAMS 79A / CLEW D34).
- Abbott Point of Care (APOC) has also determined that when testing with i-STAT cartridges that report pH, pCO2 and pO2, an i-STAT Alinity instrument with software version OSi05 (JAMS79A / CLEW D34) that is customized to correct for user entered patient temperature may report incorrect pH and/or pCO2 results.
Affected products
i-STAT Alinity Instrument with Analyzer Software
Lot or serial number
All lots.
Model or catalog number
02R60-01
Companies
- Manufacturer
-
Abbott Point of Care Inc.
100 and 200 Abbott Park Rd.
Abbott Park
60064
Illinois
UNITED STATES