Health product recall

i-STAT Alinity Instrument with Analyzer Software (2017-12-21)

Starting date:
December 21, 2017
Posting date:
January 29, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65794

Affected products

i-STAT Alinity Instrument with Analyzer Software

Reason

This product action addresses two separate issues:

  1. Abbott Point of Care (APOC) has determined that an increased frequency of quality check failure code 90-02 when testing i-STAT cartridges with the i-STAT Alinity software version Osi05 (JAMS 79A / CLEW D34).
  2. Abbott Point of Care (APOC) has also determined that when testing with i-STAT cartridges that report pH, pCO2 and pO2, an i-STAT Alinity instrument with software version OSi05 (JAMS79A / CLEW D34) that is customized to correct for user entered patient temperature may report incorrect pH and/or pCO2 results.

Affected products

i-STAT Alinity Instrument with Analyzer Software

Lot or serial number

All lots.

Model or catalog number

02R60-01

Companies
Manufacturer
Abbott Point of Care Inc.
100 and 200 Abbott Park Rd.
Abbott Park
60064
Illinois
UNITED STATES