This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

I-Stat 1 Analyzer System - PT/LNR Cartridge and Controls (2013-10-28)

Starting date:
October 28, 2013
Posting date:
December 4, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36981

Recalled products

A. I-Stat 1 Analyzer System - PT/LNR Cartridge and Controls

Reason

Abbott Point of Care has determined that I-stat PT/INR cartridges have the potential to exhibit incorrectly elevated results. Internal studies have demonstrated that I-stat PT/INR results are elevated by an average of approximately 10% as compared to the international reference preparation (RTF/09) in the therapeutic range of 1.8 to 3.0 INR.

Affected products

A. I-Stat 1 Analyzer System - PT/LNR Cartridge and Controls

Lot or serial number
  • C13139 to C13270A
  • R13149A
  • R13151
  • R13151A
Model or catalog number
  • 03P89-24
  • 04J50-01
  • 04J50-02
Companies
Manufacturer

Abbott Point of Care Inc.

100 and 200 Abbott Park Rd.

Abbott Park

60064

Illinois

UNITED STATES