I-stat 1 Analyzer system - G3+ Cartridge (2014-01-30)

Starting date:
January 30, 2014
Posting date:
March 20, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38453

Recalled Products

I-stat 1 Analyzer system - G3+ Cartridge 

Reason

Abbott point of care Inc. (APOC) has determined that I-stat G3+ cartridges from cartridge lot number N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results. Abbott point of care has determined that some individual patient results generated with this cartridge lot have the potential to exhibit incorrectly elevated pco2 and depressed pH results. Internal studies have demonstrated that PCO2 results can be elevated by 10 to 25 mm HG, pH results can be depressed by 0.10 to 0.15. All other G3+ cartridge lots are unaffected by this issue.

Affected products

I-stat 1 Analyzer system - G3+ Cartridge 

Lot or serial number

03P78-25

Model or catalog number

N13183

Companies
Manufacturer
Abbott Point of Care Inc.
100 and 200 Abbott Park Rd.
Abbott Park
60064
Illinois
UNITED STATES