I-stat 1 Analyzer system - G3+ Cartridge (2014-01-30)
- Starting date:
- January 30, 2014
- Posting date:
- March 20, 2014
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-38453
Recalled Products
I-stat 1 Analyzer system - G3+ Cartridge
Reason
Abbott point of care Inc. (APOC) has determined that I-stat G3+ cartridges from cartridge lot number N13183 have the potential to exhibit incorrectly elevated PCO2 and depressed pH results. Abbott point of care has determined that some individual patient results generated with this cartridge lot have the potential to exhibit incorrectly elevated pco2 and depressed pH results. Internal studies have demonstrated that PCO2 results can be elevated by 10 to 25 mm HG, pH results can be depressed by 0.10 to 0.15. All other G3+ cartridge lots are unaffected by this issue.
Affected products
I-stat 1 Analyzer system - G3+ Cartridge
Lot or serial number
03P78-25
Model or catalog number
N13183
Companies
- Manufacturer
-
Abbott Point of Care Inc.
100 and 200 Abbott Park Rd.
Abbott Park
60064
Illinois
UNITED STATES