This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Standard Offset Cup Impactors (2014-03-28)

Starting date:
March 28, 2014
Posting date:
September 9, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41305

Recalled Products

Standard Offset Cup Impactors

Reason

Greatbatch has determined that when a Standard Offset Cup Impactors is sterilized per the current instructions for use (IFU), the device may not be sterile. Similarly, if the device is removed from the sterilizer after being sent through steam sterilization per the IFU provided, it might not remain sterile because it is not completely dry.

Affected products

Standard Offset Cup Impactors 

Lot or serial number

Not applicable  

Model or catalog number
  • 2999-60-834
  • 2999-60-834
  • 2999-63-518
  • 2999-63-518
  • E119814
  • E119814
Companies
Manufacturer

Greatbatch Medical

4545 Kroemer Road

Fort Wayne

46818

Indiana

UNITED STATES