Health product recall

Standard Cranial Reference Array

Starting date:
August 17, 2012
Posting date:
November 26, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15910

Recalled Products

  1. Standard Cranial Reference Array

Reason

There are specific pairs of arrays that might add an inaccuracy of more than 1 mm to the registration result during the exchange due to the combination of their tolerance limits. This effect could potentially cause an inaccurate display of instruments by the navigation system in the region of interest, compared to the actual patient anatomy.

Affected products

A. Standard Cranial Reference Array

Lot or serial number

000000001085709029
9137-07-002-41735

Model or catalog number

41735

Companies
Manufacturer
BrainLab AG