STA - Liquid Anti Xa (2014-07-28)

Starting date:
July 28, 2014
Posting date:
September 3, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41243

Recalled Products

STA - Liquid Anti Xa

Reason

The Stago hotline/call center has received some customer complaints regarding a misreading of reagent 1 (substrate) barcode on STA analyzers. Diagnostica Stago has investigated and confirmed that the barcode is incorrectly coded. Consequently, when the barcode is scanned on STA analyzers, the last 3 digits displayed are "066" instead of "366". When the corresponding flyer is scanned, the STA analyzer displays an error message because the lot of the vial does not match with the lot on the flyer.

Affected products

STA - Liquid Anti Xa

Lot or serial number

112366

Model or catalog number

00311US

Companies
Manufacturer

Diagnostica Stago SAS

9, Rue Des Freres Chausson

Asnieres

92600

FRANCE