This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

STA Coagulation System - Prothrombin Time, STA-R Evolution System - Coagulation System Prothrombin Time, and STA Satelllite - Neoplastine CI (2017-10-25)

Starting date:
October 25, 2017
Posting date:
December 21, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-65496

Affected products

  1. STA Coagulation System - Prothrombin Time
  2. STA-R Evolution System - Coagulation System Prothrombin Time
  3. STA Satelllite - NEOPLASTINE CI

Reason

Following customer complaints, Stago has investigated on a potential defect of homogeneity within lot and had confirmed a performance defect on a small number of vials for some lots STA - Neoplastine CI Plus. Indeed, for these lots, some vials give longer clotting times than expected (decreased pt %). No complaints have been received from Canadian customers.

Affected products

A. STA Coagulation System - Prothrombin Time

 

Lot or serial number

250246(RU)

251308

251626

251730

251772

Model or catalog number

606

Companies
Manufacturer

Diagnostica Stago SAS

3 ALLEE THERESA

ASNIERES

92600

FRANCE



B. STA-R Evolution System - Coagulation System Prothrombin Time

 

Lot or serial number

250246(RU)

251308

251626

251730

251772

Model or catalog number

606

Companies
Manufacturer

Diagnostica Stago SAS

3 ALLEE THERESA

ASNIERES

92600

FRANCE



C. STA Satelllite - NEOPLASTINE CI

Lot or serial number

250246(RU)

251308

251626

251730

251772

Model or catalog number

606

Companies
Manufacturer

Diagnostica Stago SAS

3 ALLEE THERESA

ASNIERES

92600

FRANCE