St80i Stress Test System
Brand(s)
Last updated
Summary
Product
St80i Stress Test System
Issue
Medical devices - Software defect
What to do
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or Serial Number |
Model or catalog number |
---|---|---|
St80i Stress Test System |
More than 10 numbers, contact manufacturer. |
860343 |
Issue
When patient demographic information is entered into the system and the patient information is saved or exported to a folder, in very rare situations, the demographic information that is displayed is that of the previous patient instead of the current patient resulting in patient data mismatch. There have been complaints related to this issue, but to date there have been no reports of patient injury.
Recall start date: Dec 20, 2021
Additional information
Details
Original published date: 2022-02-01
Alert / recall type
Health product recall
Category
Health product - Medical device - Cardiovascular
Companies
Philips Medical Systems
3000 Minuteman Road, Andover, Massachusetts
United States, 01810