SpyScope DS Access and Delivery Catheter (2017-12-20)
- Starting date:
- December 20, 2017
- Posting date:
- January 29, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-65792
Affected products
SpyScope DS Access and Delivery Catheter
Reason
Potential for these devices to not perform as intended. An internal investigation found that the working channel sleeve not be adequately bonded inside of these devices. This may result in the working channel sleeve protruding from the camera cap located at the distal tip of the catheter. The most common injury reported has been minor tissue damage caused by the protruding working channel sleeve scraping along the duct walls during the procedure.
Affected products
SpyScope DS Access and Delivery Catheter
Lot or serial number
- 21159396
- 21169977
- 21212151
- 21237929
- 21264826
- 21332229
Model or catalog number
M00546600