Spotcheck - Neonatal Galt Microplate Reagent Kit
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Spotcheck - Neonatal Galt Microplate Reagent Kit |
231011 |
81-4061-00K |
Issue
On Nov. 28, 2023, a customer notified manufacturer that the NADH stock standard included in the Neonatal Galt Microplate Reagent Kit was not working as expected. Manufacturer determined that NADH stock standard, lot 230813, was manufactured with the incorrect fill quantity. Neonatal Galt Microplate Reagent Kit, lot 231011, contains the defective NADH stock standard regent, lot 230813.
Recall start date: November 28, 2023
Additional information
Details
Astoria-Pacific, Inc.
15130 Se 82nd Drive, Clackamas, Oregon, United States, 97015
Get notified
Receive emails about new and updated recall and safety alerts.