Health product recall

Spinal Rod Cutter (2020-11-25)



Starting date:
November 25, 2020
Posting date:
December 4, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74473


Last updated:

Affected Products

Spinal Rod Cutter

Reason

Zimmer Biomet is conducting a medical device recall for all lots of 00-3925-002-00 Spinal Rod Cutter maximum pin diameter 1/4 inch (6.4M) due to the potential for fracture during use. The cutter is primarily used in spine procedures to cut stainless steel rods. If the pin cutter were to fracture during use, it would be easily recognized. The highest severity event may result if the cutters were to fracture intra-operatively in an internal fixation procedure and a fragment creates a puncture wound that resulted in permanent impairment of a body function or damage to a body structure.

Affected products

Spinal Rod Cutter

Lot or serial number

All lots.

Model or catalog number

392500200

Companies
Manufacturer

Zimmer Inc.

1800 West Center Street

Warsaw

46580

UNITED STATES