Health product recall

Spectra Optia Apheresis (2019-05-21)

Starting date:
May 21, 2019
Posting date:
June 14, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70205



Last updated: 2019-06-14

Affected Products

Spectra Optia Apheresis

Reason

A connection problem to the closed circuit card assembly (CCA) of the Gen2 ebox (component of the Spectra Optia Apheresis System) was discovered. The connection issue potentially affects multiple equipment operations which may impact a planned or ongoing treatment procedures. Investigation showed that a CCA was not seated properly on it's physical supports (pems) due to the pems being various heights. This variation allows for potential bowing of the CCA,thus causing connection issues.

Affected products

Spectra Optia Apheresis

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

61000

Companies
Manufacturer

Terumo BCT, Inc.