Spectra Optia Apheresis (2019-05-21)
- Starting date:
- May 21, 2019
- Posting date:
- June 14, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70205
Last updated:
Affected Products
Spectra Optia Apheresis
Reason
A connection problem to the closed circuit card assembly (CCA) of the Gen2 ebox (component of the Spectra Optia Apheresis System) was discovered. The connection issue potentially affects multiple equipment operations which may impact a planned or ongoing treatment procedures. Investigation showed that a CCA was not seated properly on it's physical supports (pems) due to the pems being various heights. This variation allows for potential bowing of the CCA,thus causing connection issues.
Affected products
Spectra Optia Apheresis
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
61000
Companies
- Manufacturer
-
Terumo BCT, Inc.
10811 W. COLLINS AVENUE
LAKEWOOD
80215
Colorado
UNITED STATES