Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
DVI Cable Used With Medical Grade Flat Panel Display With Touchscreen |
012-089X-XX (cable) |
94267 |
DVI Cables Used With Qube Compact Monitor |
012-089X-XX (cable) |
91390 |
DVI Cable Used With Bedside Monitor |
012-089X-XX (cable) |
91393 |
Issue
While conducting 60601-1-2 testing for ongoing product development, it was found that the display cables used to connect an Xprezzon monitor(91393) or Qube compact monitor(91390) to an Xprezzon display(94267) radiates emissions outside the approved limits.