Health product recall

Spacelabs display cables for Xprezzon® or Qube Bedside Monitors

Last updated

Summary

Product
Spacelabs display cables for Xprezzon® or Qube Bedside Monitors
Issue
Medical devices - Device compatibility
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

DVI Cable Used With Medical Grade Flat Panel Display With Touchscreen

012-089X-XX (cable)

94267

DVI Cables Used With Qube Compact Monitor

012-089X-XX (cable)

91390

DVI Cable Used With Bedside Monitor

012-089X-XX (cable)

91393

Issue

While conducting 60601-1-2 testing for ongoing product development, it was found that the display cables used to connect an Xprezzon monitor(91393) or Qube compact monitor(91390) to an Xprezzon display(94267) radiates emissions outside the approved limits.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Cardiovascular
Companies

Spacelabs Healthcare Inc.

35301 Se Center Street, Snoqualmie, Washington, United States, 98065

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78311

Get notified