This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Southmedic Oxygen Masks (2017-02-15)

Starting date:
February 15, 2017
Posting date:
March 24, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62776

Affected Products

A. Ophthalmic Capnoxygen Mask

B. OxyArm-OxyMask Model - RTM

C. OxyArm-OxyPlus

Reason

On February 14, Southmedic received a customer complaint indicating a potential occlusion of the CO2 sampling line on a Southmedic Capnography Mask. Inspection of current stock of the lot number in question did not identify any issues however investigation into additional lots manufactured with the same subassembly resulted in identification of occluded CO2 sampling lines.  Subsequently all stock of various product stock keeping units (SKUs) manufactured with a CO2 sampling line were placed on hold for further inspection.

Affected products

A. Ophthalmic Capnoxygen Mask

Lot or serial number

W35906

W35983

Model or catalog number

CO2-01-SLM

CO2-03-SLM

Companies
Manufacturer

Southmedic Inc.

50 Alliance Blvd

Barrie

L4M 5K3

Ontario

CANADA



B. OxyArm-OxyMask Model - RTM

Lot or serial number

W35266

Model or catalog number

OM-2125-8

Companies
Manufacturer

Southmedic Inc.

50 Alliance Blvd

Barrie

L4M 5K3

Ontario

CANADA



C. OxyArm-OxyPlus

Lot or serial number

W35909

Model or catalog number

OP-2125-8

Companies
Manufacturer

Southmedic Inc.

50 Alliance Blvd

Barrie

L4M 5K3

Ontario