Health product recall

SORIN HEATER COOLER SYSTEM 3T (2015-06-12)

Starting date:
June 12, 2015
Posting date:
July 8, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54106

Recalled Product

A. SORIN HEATER COOLER SYSTEM 3T

Reason

Sorin has become aware that the actual disinfection practices and water maintenance by some users has not been conducted according to the IFU. Without this, organisms like mycobacteria can multiply in the heater cooler device and potentially form biofilm. Fluid leakage or aerosolization generated by a contaminated water circuit during device operation could result in patient contact and a contaminated surgical site.

Affected products

A.SORIN HEATER COOLER SYSTEM 3T

Lot or serial number

>100 NUMBERS, CONTACT MFR

Model or catalog number
  • 16-02-81
  • 16-02-82
  • 16-02-85
Companies
Manufacturer
Sorin Group Deutschland GMBH