Health product recall

Sorin Centrifugal Pump CP5 System and Control Panel (2017-10-27)

Starting date:
October 27, 2017
Posting date:
November 10, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-65082

Affected products

A. SORIN CENTRIFUGAL PUMP CP5 SYSTEM
B. SORIN CENTRIFUGAL PUMP CP5 PUMP CONTROL PANEL

Reason

During internal testing, LivaNova identified a possibility that automatic line closure by The Electrical Remote-Controlled Tubing Clamp (ERC) after detection of an air bubble could take longer than is intended by design in the CP5 system. This could lead to the remote possibility of an air bubble reaching a patient before the clamp closes if the blood flow rate is high and the distance between the bubble sensor and the patient is set at the minimum required by the IFU (1 meter).

Affected products

A. SORIN CENTRIFUGAL PUMP CP5 SYSTEM

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

60-00-60

Companies
Manufacturer
LIVANOVA DEUTSCHLAND GMBH
LINBERGHSTRASSE 25
MUNICH
80939
GERMANY

B. SORIN CENTRIFUGAL PUMP CP5 PUMP CONTROL PANEL

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

60-02-60

Companies
Manufacturer
LIVANOVA DEUTSCHLAND GMBH
LINBERGHSTRASSE 25
MUNICH
80939
GERMANY