This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

SOFTWARE FOR LIFECYCLE FOR PRENATAL SCREENING (2016-04-01)

Starting date:
April 1, 2016
Posting date:
April 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58026

Affected products 

A. SOFTWARE FOR LIFECYCLE FOR PRENATAL SCREENING

Reason

Manufacturer became aware that the latest software version has an incorrectly defined calculation parameter in one of the proportion equations for ductus venosus pulsatility index (DVPI).

Affected products

A. SOFTWARE FOR LIFECYCLE FOR PRENATAL SCREENING

Lot or serial number

v4.0

v4.0 Rev. 2

v4.0 Rev. 3

v4.0 Rev. 4

Model or catalog number

5014-0020

Companies
Manufacturer

Wallac Oy

Mustionkatu 6, P.O. Box 10

Turku

20750

FINLAND