This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Smiths Emdical Blood Sampling Kits (2013-11-18)

Starting date:
November 18, 2013
Posting date:
December 17, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37217

Recalled products

  1. Pro-Vent Plus Arterial Blood Sampling Kit
  2. Dry Line Draw Plus Arterial Blood Sampling Kit
  3. Umbilical Line Draw Plus Arterial Blood Sampling Kit

Reason

Smith Medical has identified an issue with ABL syringes when used with Radiometer brand blood gas analyzers. In August 2013, Smiths Medical began receiving a small number of complaints regarding use of the ABL syringes with Radiometer blood analyzers. The plunger tip in the ABL syringe does not remain stationary when the probe of the analyzer extends into the syringe. This allows air to leak into the sample at the luer, stopping the blood from entering the analyzer probe and preventing the analysis of blood. This can result in difficulties aspirating the sample or the plunger may get pushed out of the syringe.  Smiths Medical has received no reports of injury associated with this issue.

Affected products

A. Pro-Vent Plus Arterial Blood Sampling Kit
 

Lot or serial number

More than 1000 numbers, contact manufacturer

Model or catalog number
  • 4042LH
  • G1777
Companies
Manufacturer
Smiths Medical ASD Inc.
10 Bowman Drive
Keene
03431
New Hampshire
UNITED STATES

B. Dry Line Draw Plus Arterial Blood Sampling Kit
 

Lot or serial number

More than 1000 numbers, contact manufacturer

Model or catalog number
  • 4042-2
  • 4043-2
  • G1762
Companies
Manufacturer
Smiths Medical ASD Inc.
10 Bowman Drive
Keene
03431
New Hampshire
UNITED STATES

C. Umbilical Line Draw Plus Arterial Blood Sampling Kit

Lot or serial number

More than 1000 numbers, contact manufacturer

Model or catalog number
  • 4040-2
Companies
Manufacturer
Smiths Medical ASD Inc.
10 Bowman Drive
Keene
03431
New Hampshire
UNITED STATES