This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

SIGNA PIONEER (2016-03-03)

Starting date:
March 3, 2016
Posting date:
March 17, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57568

Affected Products     

A. SIGNA PIONEER

Reason

GE Healthcare has recently become aware of a potential safety issue with a missing ferrous warning label on SIGNA Pioneer table control box (TCB). Ferrous material in the TCB could cause unexpected attraction to the MR system's high-strength magnetic field during servicing activities. This could create a potential risk of serious injury to service personnel who attempt to inappropriately remove the TCB from the magnetic field. There have been no events reported as a result of this issue. Clinical scanning of patients is not affected by this safety notification.

Affected products

A. SIGNA PIONEER

Lot or serial number

001719TA4

Model or catalog number

SIGNA PIONEER

Companies
Manufacturer

GE Healthcare Japan Corporation

7-127 Asahigaoka, 4 Chome, Hinoshi

Tokyo

191-8503

JAPAN