Signa, Discovery and Optima Products
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
Discovery Mr750w 3.0t Magnetic Resonance Imaging System - Main Unit |
613737MR1, 506648MR4 |
DISCOVERY MR750W 3.0T |
Signa Architect |
403777SIGNA3T, 905527MR1, 604582MR750, E416249MR2, 780699MR3T |
SIGNA ARCHITECT |
Discovery Mr750 Mr System - Main Unit |
More than 10 numbers, contact manufacturer. |
DISCOVERY MR750 3.0T |
Signa Hdxt 1.5t Mr System - Main Unit |
More than 10 numbers, contact manufacturer. |
1.5T SIGNA HDXT MR SYSTEM |
Optima Mr450w Mri System - Main Unit - Gem |
More than 10 numbers, contact manufacturer. |
M7000YN |
Signa Pet/Mr |
604822MRP1 |
SIGNA PET/MR |
Signa Pioneer |
403253MR1, 250598MR1 |
SIGNA PIONEER |
Signa 1.5t Excite Hd - Magnet |
519663MRC, 416813MR3, 416369MR2, 418653MRI |
SIGNA 1.5T EXCITE HD MR SYSTEM |
Signa Hdxt 3.0t Mr System - Main Unit |
604739MR, 416369MR3T, 604270MR3T |
SIGNA HDXT 3.0T MR SYSTEM |
Signa Premier |
514934MR4 |
SIGNA PREMIER |
1.5t Signa Horizon Lx Magnetic Resonance System |
819333MRI, 778432MOBMR, 519663MRC1, 819825MRI |
M1000RA/RD/RF/RG |
Discovery Mr450 1.5t Mr System - Main Unit |
250727MR2 |
DISCOVERY MR450 1.5T MR SYSTEM |
Optima Mr450w Mri System - Main Unit - 16ch |
905522MR2, 519685MR2, 416212MR450, 902893MR1, 450466MR1 |
OPTIMA MR450W 1.5T |
Optima Mr450w Mri System - Main Unit - 32ch |
613938MR1, 780532MR1, 403670MR1 |
M7004ZJ |
Issue
GE Healthcare (GEHC) MR systems could potentially have a cryogen ventilation system that does not meet the venting requirements. Failure to have proper venting could present a safety issue if the cryogen gas is vented into the MR room during a magnet quench, potentially depriving the room of oxygen.
Recall start date: Jan 4, 2022
Additional information
Details
Ge Medical Systems LLC
3200 North Grandview Blvd, Waukesha, Wisconsin
United States, 53188