Health product recall

SIGNA ARTIST (2021-08-30)



Starting date:
August 30, 2021
Posting date:
September 17, 2021
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-76439


Last updated:

Affected Products

SIGNA ARTIST

Reason

GE healthcare has recently become aware of an issue that impacts breast imaging when running the Sagittal vibrant application with asset. If the Soslabwrap factor is larger than the default value, some slices may be missing, which can lead to a gap of anatomy in the 3D volume images.

Affected products

SIGNA ARTIST

Lot or serial number

1- PG45S2000027SC

2- PG45S2000041SC

Model or catalog number

SIGNA ARTIST

Companies
Manufacturer

GE HEALTHCARE (TIANJIN) COMPANY LIMITED

NO. 266 JINGSAN ROAD

TIANJIN

300308

CHINA