Health product recall

SICAT Implant V2.0 (2020-07-14)



Starting date:
July 14, 2020
Posting date:
August 14, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-73683


Last updated:

Affected Products

SICAT Implant V2.0

Reason

SICAT IMPLANT is a standalone software application, i.e. a radiological visualization software for diagnosis and dental implant planning.

We found that implant planning data exported from SICAT IMPLANT are not correct for one specific export format.

The error in the export format might cause dentists to manufacture (or have manufactured) surgical guides which are incorrect.

Using those guides might cause drilling into the jaw to be either not deep enough or to deep.

A dentist might have to stop a surgical procedure, or the drilling when to deep might cause damage to structures within the jaw.

Affected products

SICAT Implant V2.0
Lot or serial number

Sicat Implant V2.0 software

Model or catalog number

SICAT IMPLANT

Companies
Manufacturer

SICAT GmbH & Co. KG.

BRUNNENALLEE 6

BONN

53177

GERMANY