This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Shockpulse Lithotropsy Generator

Starting date:
November 30, 2016
Posting date:
January 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61568

Reason

Revised labeling- the rating label on the back of the ultrasonic generator specifies a voltage range of 100V to 240V , and based on tests this needs to be revised to 100V to 150V.

Affected products

Shockpulse Lithotropsy Generator

Lot or serial number

CG0032

CG0059

CG0129

CG0131

CG0158

Model or catalog number

SPL-G

Companies

Manufacturer

Cybersonics Inc.

5340 Fryling Road, Suite 101

Erie

16510

Pennsylvania

UNITED STATES