This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Shaver Handpiece Sterilization Tray (2014-05-30)

Starting date:
May 30, 2014
Posting date:
July 3, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40217

Recalled Products

Shaver Handpiece Sterilization Tray

Reason

During testing shaver hand piece tray failed Sterrad 100s half cycle testing and did not demonstrate a sterility assurance level of 10-6 for the TPS shaver. It has been determined that the reprocessing guide (P10299 Rev. A) is the non-conforming part of the shaver hand piece tray and the sterilization method with inadequate sterilization validation (Sterrad 100s) needs to be removed as a sterilization option for the TPS small joint shaver.

Affected products

Shaver Handpiece Sterilization Tray

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

0272-700-000

Companies
Manufacturer

Stryker Endoscopy

5900 Optical Court

San Jose

95138

California

UNITED STATES