Health product recall

Shaver Handpiece Sterilization Tray (2014-05-30)

Starting date:
May 30, 2014
Posting date:
July 3, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-40217

Recalled Products

Shaver Handpiece Sterilization Tray

Reason

During testing shaver hand piece tray failed Sterrad 100s half cycle testing and did not demonstrate a sterility assurance level of 10-6 for the TPS shaver. It has been determined that the reprocessing guide (P10299 Rev. A) is the non-conforming part of the shaver hand piece tray and the sterilization method with inadequate sterilization validation (Sterrad 100s) needs to be removed as a sterilization option for the TPS small joint shaver.

Affected products

Shaver Handpiece Sterilization Tray

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

0272-700-000

Companies
Manufacturer
Stryker Endoscopy
5900 Optical Court
San Jose
95138
California
UNITED STATES