SERVO-I Ventilator System - CO2 Analyzer (2015-02-09)

Starting date:
February 9, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-43835

Recalled Products 

  1. SERVO-I Ventilator System - CO2 Analyzer

Reason

The affected CO2 analyzer modules may experience a problem at the end of the calibration of the Capnostat 3 Sensor, causing the calibration to fail and activate the medium priority alarm "CO2 module error" on the user interface of the Servo Ventilator. This issue only affects the optional CO2 analyzer functionality, preventing it from displaying any measured CO2 values. It has no effect on any other functionality of the ventilator.

Affected products

A. SERVO-I Ventilator System - CO2 Analyzer

Lot or serial number

More than 10 lots, please contact the manufacturer.

Model or catalog number
  • 65 23 588 E407E
Companies
Manufacturer

Maquet Critical Care AB

Rontgenvagen 2

Solna