Servo-N
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
---|---|---|
Servo-N |
v. 4.4 or 4.5 with HFO-option |
6688600 |
Issue
A limitation of the intended performance of Servo-N during high frequency oscillatory ventilation (HFOV) when using the Fisher & Paykel Neonatal Circuit (950N81*) and the humidifier FP950. The use of a relatively small endotracheal (ET) tube in relation to body weight and in combination with the Fisher & Paykel Neonatal Circuit* that is used with the FP950 humidifier limits the delivered Pampl and therefore reduces the delivered VThf.
Recall Start Date: January 22, 2024
Additional information
Details
Maquet Critical Care Ab
Rontgenvagen 2, Solna, Sweden, 17154