This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
A-Series (A5/A7) Anesthesia Delivery System
- Starting date:
- July 14, 2017
- Posting date:
- September 5, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-64346
Reason
Mindray has become aware of the potential for a software anomaly to manifest on the a-series anesthesia system. No adverse patient event was reported. This potential issue is related to the system's electrically erasable programmable read-only memory (EEPROM) and will only occur if the system's total number of write cycles to a particular range of memory locations has exceeded the life specification of the EEPROM. The issue will only manifest at start up or at the end of a case. It will not occur during a case. Two scenarios may occur resulting from this issue.
1) The unit will not be able to enter standby mode when the user confirms "discharge" (in software bundle version < 02.06.00) or "end case" (in software bundle version > 02.06.00) in the dialog box. Under these conditions, the unit can only enter the standby mode by manually unchecking "restore default settings" in the same dialog box. When the "restore default settings" in the dialog box is unchecked, the previous settings will be applied to the next case instead of the user established default settings. If left unaddressed for a period of time, this second scenario will be encountered.
2) At startup the unit will automatically skip over the startup leak test, as well as standby mode and immediately enter either manual or automatic ventilation mode depending on the position of the manual/auto switch. This potential issue affects A-series systems with software bundle version 02.02.00 through 02.11.00 and some units with 02.11.01.
Affected products
A-Series (A5/A7) Anesthesia Delivery System
Lot or serial number
0.02.00-02.11.00 SOFTWARE
Model or catalog number
0631B-PA00012
Companies
- Manufacturer
-
Shenzhen Mindray Bio-Medical Electronics Co., Ltd.
Mindray Building, Suite 3B, Keji 12th Road South, Hi-Tech Industrial Park
Nanshan, Guangdong
518057
CHINA