This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

SEAC Dental System - Model 2020 (2013-11-19)

Starting date:
November 19, 2013
Posting date:
January 6, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37253

Recalled products

  1. SEAC Dental System - Model 2020

Reason

Field correction to the unit. Addition of a 100 PSI pressure relief valve and service warning label.  Operational pressure in excess of 100 PSI can result in rupture of the air reservoir tank.

Affected products

A. SEAC Dental System - Model 2020

Lot or serial number
  • 50341 to 592066
Model or catalog number
  • 2020
Companies
Manufacturer

ASI Medical, Inc.

130 - 13792 Compark Blvd.

Englewood

80112

Colorado

UNITED STATES