Health product recall

Sapphire Epidural Set Yellow-Striped Microbore (2018-11-04)

Starting date:
November 4, 2018
Posting date:
November 28, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-68438

Affected products

Sapphire Epidural Set Yellow-Striped Microbore

Reason

ICU Medical has identified an issue with the yellow striped sapphire epidural set. It was identified that the inclusion of both a 0.2 micron filter and a Pressure Activated Valve (PAV) with the filter in a proximal location can increase the probability of distal occlusions alarms. This configuration is limited to this sapphire epidural set.

Affected products

Sapphire Epidural Set Yellow-Striped Microbore

Lot or serial number

All lots.

Model or catalog number
  • 16373-02
Companies
Manufacturer
ICU Medical Inc.
600 NORTH FIELD DRIVE
LAKE FOREST
60045
Illinois
UNITED STATES