Sapphire Epidural Set Yellow-Striped Microbore (2018-11-04)
- Starting date:
- November 4, 2018
- Posting date:
- November 28, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68438
Affected products
Sapphire Epidural Set Yellow-Striped Microbore
Reason
ICU Medical has identified an issue with the yellow striped sapphire epidural set. It was identified that the inclusion of both a 0.2 micron filter and a Pressure Activated Valve (PAV) with the filter in a proximal location can increase the probability of distal occlusions alarms. This configuration is limited to this sapphire epidural set.
Affected products
Sapphire Epidural Set Yellow-Striped Microbore
Lot or serial number
All lots.
Model or catalog number
- 16373-02
Companies
- Manufacturer
-
ICU Medical Inc.