Health product recall

Salmonella Agglutinating Serum O poly A-S (2016-06-10)

Starting date:
June 10, 2016
Posting date:
June 24, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Hospitals, General Public, Healthcare Professionals
Identification number:
RA-59034

Affected products

A. Salmonella Agglutinating Serum O poly A-S

Reason

The overall performance of the antisera has deteriorated and this is more marked in some factors than others. The product was deemed not fit for purpose as it had deteriorated since being put on the market. It was still agglutinating with most salmonella cultures tested against, but was slower against some of the higher somatics, as it is not detecting within one minute which is stated in the instructions for use.

Affected products

A. Salmonella Agglutinating Serum O poly A-S

Lot or serial number

1335

Model or catalog number

PL 6002

Companies
Manufacturer

Pro-Lab Diagnostics

3 Bassendale Road

Bromborough, Merseyside

CH62 3QL

UNITED KINGDOM