Health product recall

Salivary Testosterone ELISA (2019-05-20)

Starting date:
May 20, 2019
Posting date:
June 14, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70197


Last updated:

Affected Products

Salivary Testosterone ELISA

Reason

The user of this assay may determine higher percentage of samples with 0 pg/mL testosterone or too low results. Saliva samples in the low measuring range of the kit (less than 50 pg/ml) were assayed significantly lower. Samples between 50 and 150 pg/mL were found approx. 30 % lower. Therefore there is a possibility of inappropriate diagnosis.

Affected products

Salivary Testosterone ELISA

Lot or serial number

66K118

66K019

Model or catalog number

SLV-3013

Companies
Manufacturer

DRG INTERNATIONAL, INC.

841 MOUNTAIN AVENUE

SPRINGFIELD

07081

New Jersey

UNITED STATES