This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

RPM Respiratory Gating System

Starting date:
March 25, 2013
Posting date:
April 29, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-28929

Recalled Products

  1. RPM Respiratory Gating System

Reason

Varian has identified an anomaly with the RPM Respiratory Gating System version 1.7.5 with 3D option. While operating with the 6-dot marker block, this version of RPM gating may produce an inverted respiratory waveform in amplitude and Phase-gating sessions.

Affected products

A. RPM Respiratory Gating System

Lot or serial number
  • More than 10 numbers, contact manufacturer
Model or catalog number
  • 100012503-01
Companies
Manufacturer

Varian Medical Systems, Inc.

3100 Hansen Way

Palo Alto

94304

California

UNITED STATES