Health product recall

Roche Hitachi System 704/717/747 Analyzers (2019-12-05)



Starting date:
December 5, 2019
Posting date:
January 20, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-72113


Last updated:

Affected Products

  1. Roche Hitachi System 704/717/747 Analyzers - Class 2
  2. Cobas Integra Glucose HK Gen.3
  3. Cobas Integra Tina-Quant Albumin Gen.2
  4. Cobas C 111 Phosphorus (PHOS2)
  5. Cobas C 111 Uric Acid (UA2)
  6. Cobas C 111 A-Amylase (AMYL2)
  7. Cobas C 111 CREJ2 - Creatinine Jaffe
  8. Cobas C 111 CREP2 - Creatinine Plus
  9. CO2-L (Bicarbonate Liquid) for Cobas C111 Analyzer
  10. CA2 (Calcium Gen.2)
  11. PHOS2 (Phosphate (Inorganic) VER.2)
  12. TPUC3 (Total Protein Urine /CSF Gen.3)
  13. TPUC3 (Total Protein Urine /CSF Gen.3)
  14. CO2-L (Bicarbonate Liquid)
  15. AMY-P (Alpha-Amylase EPS Pancreatic)
  16. ETOH2 (Ethanol Gen.2)
  17. CREP2 (Creatinine Plus Ver. 2)
  18. CREJ2 (Creatinine Jaffe Gen. 2)
  19. UA2 (Uric Acid VER.2)

Reason

Following the internal study  (conducted by the manufacturer) on endogenous and drug interference which identified new endogenous interferences claims (RR201836), additional endogenous and individual drug interferences were identified, but were evaluated not likely to cause a relevant medical risk to patients and users. Therefore, method sheets for the cobas c 111, cobas c 311/501/502, cobas c 701/702 and cobas Integra 400 Plus platforms will be updated with new interference claim.  In addition, the reduced concentration of the urea interference claim for all cobas c analyzers for amyl2/amy-p is stated.

 

Affected products

A. Roche Hitachi System 704/717/747 Analyzers - Class 2

Lot or serial number

29695701

30558501

30558801

32402901

32403201

39993801

Model or catalog number

10759350190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



B. Cobas Integra Glucose HK Gen.3

Lot or serial number

37118201

38172101

39105101

39929101

40879901

42138401

43230301

44444401

Model or catalog number

4404483190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



C. Cobas Integra Tina-Quant Albumin Gen.2

Lot or serial number

32279101

34022301

34852501

36281701

37742301

39087401

40384301

41549301

42591701

44042501

Model or catalog number

4469658190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



D. Cobas C 111 Phosphorus (PHOS2)

Lot or serial number

37138201

39515401

42227901

44681301

Model or catalog number

05401780190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



E. Cobas C 111 Uric Acid (UA2)

Lot or serial number

38914401

41582001

43133201

Model or catalog number

4657608190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



F. Cobas C 111 A-Amylase (AMYL2)

Lot or serial number

38408201

40169901

42941601

44053901

Model or catalog number

05401496190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



G. Cobas C 111 CREJ2 - Creatinine Jaffe

Lot or serial number

31788801

35247501

38184001

41615101

Model or catalog number

05401755190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



H. Cobas C 111 CREP2 - Creatinine Plus

Lot or serial number

39514801

40633501

41561401

42448801

Model or catalog number

05401470190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



I. CO2-L (Bicarbonate Liquid) for Cobas C111 Analyzer

Lot or serial number

35776601

36579401

36759901

39391901

Model or catalog number

5061482190

 

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



J. CA2 (Calcium Gen.2)

Lot or serial number

36293901

37154501

40148501

40821801

42933501

Model or catalog number

5061504190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



K. PHOS2 (Phosphate (Inorganic) VER.2)

Lot or serial number

36040901

38580801

40875201

40881101

41738501

42820301

42857401

44076201

Model or catalog number

3183793122

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



L. TPUC3 (Total Protein Urine /CSF Gen.3)

Lot or serial number

36286301

38650201

40738901

43134901

Model or catalog number

3333825190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



M. TPUC3 (Total Protein Urine /CSF Gen.3)

Lot or serial number

38412101

39928101

41581501

43654501

Model or catalog number

3183742122

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



N. CO2-L (Bicarbonate Liquid)

Lot or serial number

32903701

33635801

36578401

36756301

36759201

38892301

39390701

39391601

40218901

Model or catalog number

3289923190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



O. AMY-P (Alpha-Amylase EPS Pancreatic)

Lot or serial number

38412701

40445201

41777001

43655201

Model or catalog number

20766623322

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



P. ETOH2 (Ethanol Gen.2)

Lot or serial number

42820301

Model or catalog number

3183777190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



Q. CREP2 (Creatinine Plus Ver. 2)

Lot or serial number

38876501

38981202

40558001

41817801

42052701

42284101

43470801

Model or catalog number

3263991190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



R. CREJ2 (Creatinine Jaffe Gen. 2)

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

4810716190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY



S. UA2 (Uric Acid VER.2)

Lot or serial number

38427601

41580001

43180101

Model or catalog number

3183807190

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY