This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

RHD BeadChip Kit (2016-07-08)

Starting date:
July 8, 2016
Posting date:
July 29, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-59600

Affected products

A. RHD BeadChip Kit

Reason

RHD master lot has incorrect deletion detection values. Lot-specific deletion detection values released into basis global software for RHD master lot 16-284 may result in an incorrect phenotypical variant call due to the genotype being erroneously detected as having an exon 5 deletion.

Affected products

A. RHD BeadChip Kit

Lot or serial number

16-284

Model or catalog number

800-10220-48

Companies
Manufacturer

BIOARRAY SOLUTIONS LTD

35 TECHNOLOGY DRIVE, SUITE 100,

Warren

07059

New Jersey

UNITED STATES