This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

REFLECTION 3 Acetabular System-Constrained Acetabular Liner (2015-05-08)

Starting date:
May 8, 2015
Posting date:
June 3, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53657

Recalled Products

A. REFLECTION 3 Acetabular System-Constrained Acetabular Liner

Reason

Smith & Nephew has initiated a product recall of select lots of the R3 constrained acetabular liners due to a manufacturing error. The R3 constrained outer liner locking groove diameter is 0.030" oversized and may not allow the locking ring to engage into the R3 shell.

Affected products

A. REFLECTION 3 Acetabular System-Constrained Acetabular Liner

Lot or serial number
  • 14HM08141
  • 14HM08148
  • 14HM08176
Model or catalog number
  • 71339160
  • 71339162
Companies
Manufacturer

Smith & Nephew Inc.

1450 Brooks Road

Memphis, Tennessee

38116

UNITED STATES