Raystation 4-11b
Brand(s)
Last updated
Summary
Product
Raystation 4-11b
Issue
Medical devices - Software defect
What to do
Contact the manufacturer if you require additional information.
Affected products
| Affected Products | Lot or Serial Number | Model or catalog number |
|---|---|---|
| Raystation 4-11b | version 8.1.2, version 10.0.1, version 7.0.0 | 10B (10.1), 7, 8B, 10A (10.0) |
Issue
In rare cases, the SSD (source to skin or surface distance) calculation algorithm will miss the entry point of the ROI and instead calculate the distance to the exit point, resulting in an incorrect SSD. The SSD will be incorrect in UI, Dicom export, plan report, and scripting. If the bug is triggered when an intended SSD is entered in Raystation, the intended SSD will be displayed and exported, but the actual SSD of the plan is shorter than intended.
Recall start date: Feb 11, 2022
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Raysearch Laboratories Ab (Publ)
Sveavagen 44, Stockholm
Sweden, 11134
Published by
Health Canada
Audience
General public
Health professionals
Hospitals
Recall class
Type II
Identification number
RA-63986
Get notified
Receive emails about new and updated recall and safety alerts.