RayStation (2018-11-10)
- Starting date:
- November 10, 2018
- Posting date:
- November 30, 2018
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-68498
Affected products
RayStation
Reason
This recall notice concerns an issue found with isocenter shifts when using dose tracking in RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayStation 7 and RayStation 8A. In these versions, isocenter shifts in dose tracking are incorrectly interpreted according to the DICOM patient coordinate system.
Affected products
RayStation
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
4.0.3
5
6
7
Companies
- Manufacturer
-
RaySearch Laboratories AB (PUBL)
Sveavagen 25
Stockholm
11134
SWEDEN