This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

RAYSTATION (2017-11-20)

Starting date:
November 20, 2017
Posting date:
November 30, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Hospitals, Healthcare Professionals
Identification number:
RA-65286

Affected products

RAYSTATION

Reason

This notice concerns an issue found with the electron dose calculation in RayStation 3.5, RayStation 4.0, RayStation 4.5, RayStation 4.7, RayStation 5, RayStation 6, RayPlan 1 (RayStation 4.9) and RayPlan 2 (RayStation 6). When calculating electron Monte Carlo dose with a very large number of histories, the dose calculation may be wrong.

Affected products

RAYSTATION

Lot or serial number

RayStation 4.7.2.5

RayStation 5.0.2.35

Model or catalog number

4.0.3

5

6

Companies
Manufacturer

RAYSEARCH LABORATORIES AB (PUBL)

SVEAVAGEN 44

STOCKHOLM

11134

SWEDEN