Raystation (2015-05-08)

Starting date:
May 8, 2015
Posting date:
June 15, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53793

Recalled Products

  1. Raystation

Reason

An issue was found with the dose calculation for some imported VMAT plans where controls points are defined with wide gantry angle spacing. When VMAT control points are distributed with a wider gantry angle spacing than 4 degrees, the dose calculation will not be accurate.

Affected products

A. Raystation

Lot or serial number
  • 3.5.0.16
  • 4.0.1.4
  • 4.0.3.4
Model or catalog number
  • 3.5
  • 4
Companies
Manufacturer

Raysearch Laboratories AB,

Sveavagen 25,

Stockholm

11134