This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Raystation (2014-09-17)

Starting date:
September 17, 2014
Posting date:
October 2, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41635

Recalled Products   

Raystation 

Reason

This notice concerns a problem with the computation of ROI voxel volumes for cases where the CT has variable slice spacing. The dose grid volumetric representation of ROI can be wrong. This affects all dose-volume properties for such ROIs, including DVH, dose statistics, clinical goals and constraints or objective functions. Furthermore, if material override ROIs are used, or if the external is not represented by contours, the bug can trigger an error in the 3D dose distribution.

Affected products

Raystation 

Lot or serial number

VERSION 4.0.3.4

Model or catalog number

4.0.3

Companies
Manufacturer
Raysearch Laboratories AB,
Sveavagen 25,
Stockholm
11134
SWEDEN