This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Range Check X II (2015-02-12)

Starting date:
February 12, 2015
Posting date:
March 20, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52583

Recalled products

  1. Range Check X II

Reason

Sysmex America, Inc. received complaints from Sysmex installation specialists and customers that level 2 of the five levels of white blood cell (WBC) linearity of range check XII (RCXLL) was recovering above the expected range on the POCH-1 OOI, XP-300 and KX-21 n, and KX-21 analyzers (three-part differential analyzers). Level 2 WBC recovery is expected to be approximately 0.5 +/- 0.5 x103/uL and recovery was above 1.0 x103/uL in some instances. When the recovery is above 1.0, the plotted line is not straight suggesting the analyzer is not linear through the entire wbc reportable range of 1.0 to 99.9 x103/uL for the three part differential analyzers. Exclusion of the level 2 produces a linear graph.

Affected products

A. Range Check X II

Lot or serial number
  • 4091
  • 4161
  • 4231
  • 4301
  • 5006
  • 5076
Model or catalog number
  • 130-4003-0
Companies
Manufacturer

Sysmex Corporation

1-5-1 Wakinohama-kaigandori

Chuo-ku