This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Rainbow Reusable Sensors (2016-04-22)

Starting date:
April 22, 2016
Posting date:
May 16, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-58426

Affected Products

Rainbow Reusable Sensors

Reason

Masimo has identified specific lots of sensors that were manufactured with incompatible configurations. This could result in sensors that will either not provide readings for the SpCO and SpMet parameters or inaccurate readings. The specific behavior will depend on the firmware version of the technology board in the device.  SpCO measurements, when displayed, could be incorrectly elevated throughout the measurement range.  SpMet measurements above 6% could be affected by a negative bias of 1.5%, on average.

Affected products

Rainbow Reusable Sensors

Lot or serial number
  • BOX LOT # 15JA8
  • BOX LOT 15HA6
  • Sensor Ref#-LOT# 24087-A15G760
  • Sensor Ref#-LOT# 24087-A15H754
  • Sensor Ref#-LOT# 24087-A15J726
  • Sensor Ref#-LOT# 24087-A15J824
  • Sensor Ref#/LOT# 24087-A15H754
Model or catalog number
  • 2696
Companies
Manufacturer

Masimo Corporation

40 Parker

Irvine

California

UNITED STATES