Health product recall

R100 Ventilator System (2013-11-15)

Starting date:
November 15, 2013
Posting date:
February 3, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Product Safety
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37527

Recalled Products

R100 Ventilator System

Reason

One report received of the actuation of ventilator alarm and buzzing sound after 16 hrs of patient oscillation in a hospital in Germany. The reason was a failure in the bonding of the diaphragm chamber producing leakage.

Affected products

R100 Ventilator System

Lot or serial number

Not Applicable

Model or catalog number

10010410001

Companies
Manufacturer
Metran Co., Ltd.
2-12-18, Kawaguchi
Kawaguchi-Shi
332-0015
JAPAN