This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

R100 Ventilator System (2013-11-15)

Starting date:
November 15, 2013
Posting date:
February 3, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Product Safety
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37527

Recalled Products

R100 Ventilator System

Reason

One report received of the actuation of ventilator alarm and buzzing sound after 16 hrs of patient oscillation in a hospital in Germany. The reason was a failure in the bonding of the diaphragm chamber producing leakage.

Affected products

R100 Ventilator System

Lot or serial number

Not Applicable

Model or catalog number

10010410001

Companies
Manufacturer

Metran Co., Ltd.

2-12-18, Kawaguchi

Kawaguchi-Shi

332-0015

JAPAN