Health product recall

Quickie Q7 Manual Wheelchair with Backrest Brackets

Starting date:
July 4, 2016
Posting date:
March 24, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62738

Reason

Sunrise Medical (US) LLC has observed failure of the original Backrest Bracket (2009-2013) which can fatigue over the lifetime of the product. When this occurs, there is a possibility of the bracket fracturing causing the backrest to detach from the chair frame. The root cause was found to be insufficient material in a high stress area. Under certain loading conditions these high stresses can lead to material fatigue and ultimately failure of the bracket.

Affected products

Quickie Q7 Manual Wheelchair with Backrest Brackets

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

EIR4

Companies

Manufacturer

Sunrise Medical (US) LLC

2842 Business Park Avenue

Fresno

93727

California

UNITED STATES